Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8026B: CAL ADJ /5500 DATE: 16-Jan-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 87 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 8026B instruction manual (PN 646414 June 1982). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_V8020.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J W 1.002 HEAD EQUIPMENT SETUP 1.003 DISP [32] WARNING 1.003 DISP HIGH VOLTAGE is used or exposed during the performance 1.003 DISP of this calibration. DEATH ON CONTACT may result if 1.003 DISP personnel fail to observe safety precautions. 1.004 HEAD CALIBRATION ACCESS PROCEDURE 1.005 DISP Disconnect the test leads and the battery eliminator 1.005 DISP if attached. Open the battery compartment, and 1.005 DISP disconnect the battery. Remove the three phillips-head 1.005 DISP screws from the bottom of the case. Turn the instrument 1.005 DISP face-up, and grasp the top cover at both sides of the 1.005 DISP input connectors. Then, pull the top cover from the UUT. 1.005 DISP Re-connect the battery to the battery jacks. 1.006 DISP Ambient temperature: 21C - 25C. 1.006 DISP Relative humidity: less than 80%. 1.006 DISP Allow the UUT 30 min. to stabilize. 1.006 DISP Turn the UUT on. 1.007 HEAD DC VOLTAGE CALIBRATION ADJUSTMENT 1.008 PIC 55_V8020 # 1.009 DISP [32] Connect the 5500A and the UUT as follows. # 1.009 DISP [32] 5500A NORMAL HI to 8026B V Ohm S # 1.009 DISP [32] 5500A NORMAL LO to 8026B COMMON 1.009 RNG 200mV 1.010 5500 190mV S 2W 1.011 DISP Adjust R5 for a reading of 190.0mV. 1.012 HEAD AC VOLTAGE CALIBRATION ADJUSTMENT 1.013 RNG 200mV AC 1.014 5500 190mV 100H SI S 2W 1.015 DISP Adjust the AC CAL potentiometer marked "F" (fine 1.015 DISP adjustment- R19) so that it is at mid-range. 1.015 DISP 1.015 DISP Adjust the AC CAl potentiometer marked "C" 1.015 DISP (coarse-adjust-R17) for a UUT reading of 190.0 1.015 DISP +/- counts. 1.015 DISP Then adjust R19 ("F" fine adjust) for a UUT reading 1.015 DISP of 190.0 (+/- 1/2 digit). 1.016 5500 * S 1.017 RNG 20V AC 1.018 5500 19V 7kH SI S 2W 1.019 DISP Adjust C9 for a reading of 19.00V (+/- 2 digits). 1.020 5500 * S 1.021 RNG 2V AC 1.022 5500 1.9V 10kH SI S 2W 1.023 DISP Adjust C2 for a reading of 1.900V (+/- 2 digits). 1.024 END